Hydrostatic pressure transmitter ECL is based on Trafag's proprietary thick-film-on-ceramic technology. The optionally configurable pressure ranges can be individually customized via interface tool and smartphone app.
Applications: process technology, water treatment (wastewater, gray water, drinking water), sea water, level of oils and fuel.
Applications | Shipbuilding, Process technology, Water treatment (wastewater, grey-water, drinking water), Seawater, Level of oils and fuels |
---|---|
Category | Level measurement |
Accuracy @ 25°C typ. | ± 0.3 % FS typ. Range 0 ... 0.1 to 0 ... 0.2 bar: ± 0.5 % FS typ. |
Measuring principle | Thick-film-on-ceramic |
Vibration | 20 g (40 ... 2000 Hz) 15 grms (20 ... 2000 Hz) |
NLH @ 25°C (BSL) typ. | ± 0.2 % FS typ. Range 0 ... 0.1 to 0 ... 0.2 bar: ± 0.3 % FS typ. |
Media temperature | max. -25°C ... +70°C |
Shock | 50 g / 8 ms |
Electrical connections | Cable PUR/Radox/PE |
Long-term stability 1 year typ. | ± 0.2 % FS typ. |
Ambient temperature | max. -25°C ... +70°C |
Protection | IP68 (2.0 bar/20 m) |
Output signal | 4 ... 20 mA |
TEB typ. @ -25 ... +85°C | ± 1.0 % FS typ. Range 0 ... 0.1 to 0 ... 0.2 bar: ± 2.0 % FS typ. |
Measuring range | 0 ... 0.1 to 0 ... 2.0 bar 0 ... 1.5 to 0 ... 30 psi |
Approval / conformity | DNV-GL EU RO Mutual Recognition Type Approval Certificate |
- ECL 8439: App for smart parameterization and cable protection
As with a smartphone app and the use of proven cable glands from process and chemical technology, amazing new solutions to known problems in hydrostatic fill level measurement are possible.